![Agilent Technologies ESA Series Скачать руководство пользователя страница 190](http://html1.mh-extra.com/html/agilent-technologies/esa-series/esa-series_manual_2868109190.webp)
190
Chapter 4
ESA Functional Tests
Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B, E4407B and E4408B
E7402A, E7403A, E7404A, and E7405A, Option 1DN
ESA
Functional T
ests
Tracking Generator Level Flatness:
E4402B, E4403B, E4404B, E4405B, E4407B and E4408B
E7402A, E7403A, E7404A, and E7405A, Option 1DN
Test Limits
Test Description
A calibrated power sensor is connected to the tracking generator output to measure the
power level at 50 MHz. The power meter is set for REL mode so that future power level
readings are in dB relative to the power level at 50 MHz. The tracking generator is then
stepped to several frequencies throughout its range. The output power difference relative
to the power level at 50 MHz is measured at each frequency and recorded.
Maximum
Maximum
Flatness
≤
10 MHz
−
3.5 dBm
+3.5 dBm
Flatness > 10 MHz
−
2.5 dBm
−
2.5 dBm
Item
Critical Specifications
(for this test)
Recommended
Agilent Model
Meters
Power Meter
438A or E4418A,
E4419A
RF Power Sensor
Frequency Range: 100 kHz to 3.0 GHz
8482A
Adapters
Type-N (f) to BNC
(m), 75-Ohm
1250-1534
Type-N (m to BNC (f)
(2)
1250-1476
Cables
(2) BNC, 122-cm
(48-in)
10503A
BNC, 23-cm (9-in)
10502A
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...