Instrument Messages
Instrument Messages
In
str
u
me
n
t Me
s
s
a
g
es
627 PSA only
609 ESA only
Media is not writable
A save was attempted to a read-only device or device that could not be
written to. Try a known-good disk.
ERROR/NUMBER “609
Restoration of NVRAM data”
IS ALREADY
USED FOR HAPPENING ./storage/directory/NRAM_Directory.cc
HAP_nramDataCorrected
651
Connect RF OUT to INPUT
Attempt to align the tracking generator without its output connected.
Connect the tracking generator RF OUT to the analyzer
INPUT
.
652
Connect Amptd Ref Output to Input
For Agilent Technologies E4402B, E4403B, E4404B, E4405B, E4407B, and
E4408B only:
you must connect the
AMPTD REF OUTPUT
to the analyzer
INPUT
with the appropriate cable.
653
Auto Align not available when using Calibration Defaults
The Auto Alignment system cannot be used until an
Align Now All
is
executed by pressing
System
,
Alignments
,
Align Now
,
All
. On all Agilent
Technologies ESA spectrum analyzer models except Agilent Technologies
E4401B and E4411B, you must connect the
AMPTD REF OUT
to the
INPUT
with the appropriate cable to perform this alignment.
For Agilent
Technologies E4401B and E4411B only:
disconnect any signals from the
INPUT
prior to performing this procedure.
701
Invalid printer response
In attempting to identify the printer an invalid response was received.
Check that you are using a supported printer. Be sure you are using the
proper cable and that it is securely fastened.
702
Unsupported printer
A printer which is recognized, but known to be unsupported was
identified. This printer cannot be used with the analyzer. For example, a
printer only supported by Microsoft Windows will generate this error.
703
Unknown printer
In attempting to identify the printer, a valid response was received but the
printer is not known to the analyzer. Use the
Custom
printer menu under
Print Setup
to configure the printer.
704
Printer interface error
An error occurred while trying to print. Make sure the printer is turned on
and properly connected.
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...