Chapter 3
39
Operation Verification
Option UN7 (BERT) Check
Configuring the Data
1. Press
Preset
.
2. For signal generators with Option UN7 only press:
Aux Fctn
>
BERT
>
BERT Off On
to
On
.
For signal generators with Option UN7 and Option 300 press:
Aux Fctn
>
BERT
>
Baseband BERT
>
BERT Off On
to
On
.
Configuring BERT Parameters
The following steps configure BERT measurement parameters.
1. Press
Configure BERT
>
Data
>
PN9
.
2. Press
Total Bits
to 10
MBits
.
3. Press
More (1 of 2)
>
BERT Resync Off On
to
Off
.
4. Press
Return
>
I/O Setup
>
Polarity Setup
>
Clock Polarity Pos Neg
to Neg.
5. Press
Return
>
Aux I/O Out
>
PN9 Out
.
6. Press
Return
>
Configure Trigger
>
Cycle Count 0
>
Enter
.
The following figure shows the signal generator’s front-panel display after completion of the these steps.