Chapter 3
33
Operation Verification
FM Accuracy Check
FM Accuracy Check
Connect the Test Equipment
Figure 3-4
FM Accuracy Equipment Setup
Configure the Signal Generator
1. Press
Preset
.
2. Press
FM/
Φ
M
>
FM Off On
. The
FM
annunciator is displayed.
3. Press
FM Rate
. Enter 1 using the numeric keypad and press the
kHz
terminator softkey.
4. Press
FM Dev.
Enter 100 using the numeric keypad and press the
kHz
terminator softkey.
5. Press
Amplitude.
Enter 7 using the numeric keypad and press the
dBm
terminator softkey.
6. Ensure that the
MOD ON
annunciator is displayed. If it is not, press
Mod On/Off
.
7. Press
RF On/Off
. The
RF ON
annunciator is displayed.
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to FM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.