Chapter 3
27
Operation Verification
Frequency Range and Accuracy Check
Frequency Range and Accuracy Check
The frequency range is tested by determining the frequency accuracy relative to the timebase at the
frequency limits of the signal generator.
Connect the Test Equipment
Figure 3-1
Frequency Range and Accuracy Equipment Setup
Configure the Signal Generator
1. Press
Preset
.
2. Press
Mod On/Off
. The
MOD OFF
annunciator is now displayed.
3. Press
Amplitude
. Enter 0 using the numeric keypad and press the
dBm
terminator softkey.
4. Press
RF On/Off
. The
RF ON
annunciator is now displayed.
Configure the Frequency Counter
1. Observe the display. Verify that the counter is frequency-locked to the 10 MHz external reference
(±1 Hz).
2. For frequencies < 150 MHz, use Channel 3. Press Freq Ratio until
CH3:
is displayed.
The input impedance of this channel is a persistent 50
Ω
.