Appendix Service Information
Calibration Record for Your Power Supply
202
Calibration Record for Your Power Supply
Step
Calibration Description
Measurement
Mode (DVM)
Supply Being
Adjusted
1
Turn on the calibration mode by holding down the "Calibrate" key as you
turn on the power supply until you hear a long beep.
2
Unsecure the power supply if secured. (See page 67)
3
Press "Calibrate" key to move down menu to voltage calibration menu.
A "
VOLT1 CAL
" is displayed. Press "Calibrate" key to select the low
voltage calibration point.
Voltage
Calibration
4
A low voltage calibration point is displayed. Enter the DVM reading by
using the knob and resolution keys. Press "Calibrate" key to save the
changes and select the middle calibration point.
V
Low voltage
calibration
5
A middle voltage calibration point is displayed. Enter the DVM reading
by using the knob and resolution selection keys. Press "Calibrate" key to
save the changes and select the high calibration point.
V
Middle voltage
calibration
6
A high voltage calibration point is displayed. Enter the DVM reading by
using the knob and resolution selection keys. Press "Calibrate" key to
save the changes and select the OVP calibration.
V
High voltage
calibration
7
"
OVP1 CAL
" is displayed. Press "Calibrate" key to run the overvoltage
calibration. A "
CALIBRATING
" is displayed to indicate that the
calibration is progressing.
V
OVP calibration
8
Connect shunt resistor (0.01
Ω
) across the output terminals. And press
"Calibrate" key to select the current calibration.
Current calibration
9
A low current calibration point is displayed. Enter the computed value
(DVM reading / by shunt resistance) by using the knob and resolution
keys. Press "Calibrate" key to save the changes and select the middle
calibration point.
A
Low current
calibration
10
A middle current calibration point is displayed. Enter the computed value
(DVM reading / by shunt resistance) by using the knob and resolution
keys. Press "Calibrate" key to save the changes and select the high
calibration point.
A
Middle current
calibration
11
A high current calibration point is displayed. Enter the computed value
(DVM reading / by shunt resistance) by using the knob and resolution
keys. Press "Calibrate" key to save the changes
A
High current
calibration
13
Repeat the steps (3) through (16) for the output2 calibration.
14
Turn off the "Power" switch to exit the calibration menu.
Exit CAL MODE
Содержание E3646A
Страница 2: ...2...
Страница 16: ...16 Contents Contents...
Страница 17: ...1 Quick Start...
Страница 27: ...2 General Information...
Страница 39: ...3 Front Panel Operation and Features...
Страница 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Страница 73: ...4 Remote Interface Reference...
Страница 125: ...5 Error Messages...
Страница 136: ...Chapter 5 Error Messages Calibration Errors 136...
Страница 137: ...6 Application Programs...
Страница 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Страница 149: ...7 Tutorial...
Страница 160: ...Chapter 7 Tutorial Remote Programming 160...
Страница 161: ...8 Specifications...
Страница 168: ...Chapter 8 Specifications Supplemental Characteristics 168 Figure 8 2 Dimensions for Rack mounting...
Страница 169: ...Appendix Service Information...
Страница 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Страница 210: ......