Chapter 5 Error Messages
Execution Errors
132
800
Outputs coupled by track system
This error occurs when coupling the outputs if the track mode is already enabled.
Tracking mode should be disabled first when coupling the outputs.
801
Outputs coupled by trigger subsystem
This error occurs when enabling the track mode if the outputs are already coupled.
Outputs should be uncoupled first to enable the power supply to operate in the track
mode.