E1463A Command Reference 49
Chapter 3
Parameters
ParameterTypes.
The following table contains explanations and examples of
parameter types you might see later in this chapter.
Linking Commands
Linking IEEE 488.2 Common Commands with SCPI Commands.
Use a
semicolon (;) between the commands. For example, *RST;*RCL 1 or
CLOS (@101);*SAV 1
Linking Multiple SCPI Commands.
Use both a semicolon (;) and a colon (:)
between the commands, such as CLOS (@101);:CLOS? (@101).
Linking Subsystem Commands
. SCPI also allows several commands within
the same subsystem to be linked with a semicolon, such as ROUT:CLOS
(@101);:ROUT:CLOS? (@101) or ROUT:CLOS (@101);CLOS? (@101).
SCPI Command Reference
This section describes the Standard Commands for Programmable
Instruments (SCPI) commands for the E1463A. Commands are listed
alphabetically by subsystem and within each subsystem.
Type
Explanations and Examples
Boolean
Represents a single binary condition that is either true or
false (ON, OFF, 1.0). Any non-zero value is considered
true.
Discrete
Selects from a finite number of values. These parameters
use mnemonics to represent each valid setting. An
example is the TRIGger:SOURce <source> command
where <source> can be BUS, EXTernal, HOLD,
IMMediate, or TTLTrgn.
Numeric
Commonly used decimal representations of numbers
including optional signs, decimal points, and scientific
notation. Examples are 123, 123E2, -123, -1.23E2, .123,
1.23E-2, 1.23000E-01. Special cases include MINimum,
MAXimum, DEFault and INFinity.
Optional
Parameters shown within square brackets ([ ]) are optional
parameters. (The brackets are not part of the command
and are not sent to the instrument.) If you do not specify a
value for an optional parameter, the instrument chooses a
default value.
For example, consider the ARM:COUNt? [<MIN | MAX>]
command. If you send the command without specifying a
parameter, the present ARM:COUNt value is returned. If
you send the MIN parameter, the command returns the
minimum count available. If you send the MAX parameter,
the command returns the maximum count available. Be
sure to place a space between the command and the
parameter.