Agilent E1330B Digital I/O Module Command Reference
59
Chapter 5
Parameters
Parameter Types. The following table contains explanations and examples of
parameter types you might see later in this chapter.
Optional Parameters. Parameters shown within square brackets (
[]
) are optional
parameters. (Note that the brackets are not part of the command and are not sent to
the instrument.) If you do not specify a value for an optional parameter, the
instrument chooses a default value. For example, consider the
DISPlay:MONitor:PORT? [<MIN| MAX| DEF>]
command. If you send the command
without specifying a parameter, the command returns the state of the port last
addressed. If you send the
MIN
parameter or the
DEF
parameter, the command
returns
0
. If you send the
MAX
parameter, the command returns
3
. Be sure to place
a space between the command and the parameter.
Keyword
Substitutions
Some commands indicate a keyword substitution by showing the keyword bold type
with an all lower case keyword. For example, in the SCPI command
MEASure:DIGital:DATA
n
[:type]
the keyword
[:type]
should be replaced by one
these parameters:
:BYTE
:WORD
:LWORd
Linking
Commands
Linking IEEE 488.2 Common Commands with SCPI Commands.
Use a semicolon between the commands. For example:
*RST;DIG:CONT2 1
or
DIG:CONT2:POL POS;*OPC?
Linking Multiple SCPI Commands.
Use both a semicolon and a colon between the commands. For example:
DIG:DATA2:POL NEG;:DIG:DATA2:BIT3 1
Parameter
Type
Explanations and Examples
Numeric
Accepts all commonly used decimal representations of numbers
including optional signs, decimal points, and scientific notation.
123, 123E2, -123, -1.23E2, .123, 1.23E-2, 1.23000E-01. Special
cases include MIN, MAX, and DEF. MIN selects minimum value
available, MAX selects maximum value available, and DEF
selects default or reset value.
Boolean
Represents a single binary condition that is either true or false.
1 or ON; 0 or OFF.
Discrete
Selects from a finite number of values. These parameters use
mnemonics to represent each valid setting.
An example is the DIGital:CONTrol
n
:POLarity <
polarity
>
command where
polarity
can be either POS or NEG.
Содержание E1330B
Страница 2: ......
Страница 10: ...8 Notes ...
Страница 11: ...9 Notes ...
Страница 12: ...10 Notes ...
Страница 25: ...Configuring the Agilent E1330B Digital I O Module 23 Chapter 2 Figure 2 6 J1 and J2 Connector Pinouts ...
Страница 32: ...30 Configuring the Agilent E1330B Digital I O Chapter 2 Notes ...
Страница 42: ...40 Using the Agilent E1330B Digital I O Module Chapter 3 Notes ...
Страница 58: ...56 Understanding the Agilent E1330B Digital I O Module Chapter 4 Notes ...
Страница 104: ...102 Agilent E1330B Digital I O Module Command Reference Chapter 5 Notes ...
Страница 146: ...144 Error Messages Appendix C Notes ...