15
Agilent DNA 7500 and DNA 12000 Assay Protocol
5
Loading the Marker
5
Press the plunger of the syringe down until it is
held by the clip.
6
Wait for exactly 30 seconds and then release the
plunger with the clip release mechanism.
7
Visually inspect that the plunger moves back at
least to the 0.3 ml mark.
8
Wait for 5 seconds, then slowly pull back the
plunger to the 1 ml position.
9
Open the chip priming station.
10
Pipette 9.0 μl of the gel-dye mix in each of the
wells marked .
Loading the Marker
1
Pipette 5 μl of green-capped marker (green )
into the well marked with the ladder symbol
and into each of the 12 sample wells.
SUHVVXUL]H
OJHOG\H
N O T E
Protect the gel-dye mix from light. Store the gel-dye mix at 4 °C when not in use for more
than 1 hour.
OPDUNHU
N O T E
Do not leave any wells empty, or the chip will not run properly. Add 5 µl of green-capped
DNA marker (green ) plus 1 µl of deionized water to each unused sample well.