
16
Agilent E1412A Multimeter Module Setup
Setting the Module Address Switch
The logical address switch factory setting is 24. Valid addresses are from
1 to 254 for static configuration (the address you set on the switch) and
address 255 for dynamic configuration. The E1412A supports dynamic
configuration of the address. This means the address is set programmatically
by the resource manager when it encounters a module with address 255 that
supports dynamic configuration.
If you install more than one multimeter, each module must have a different
logical address. If you use a VXIbus command module, the logical address
must be a multiple of eight (e.g., 32, 40, 48, etc.) Each instrument must have
a unique secondary address which is the logical address divided by eight.
Note
When using an E1405A/B or E1406A as the VXIbus resource manager
with SCPI commands, the multimeter's address switch value must be a
multiple of 8.
Figure 1-1. Setting the Logical Address
Содержание Agilent E1412A
Страница 2: ......
Страница 10: ...8 Contents Notes...
Страница 14: ...12 Notes...
Страница 15: ...13 Notes...
Страница 16: ...14 Notes...
Страница 20: ...18 Agilent E1412A Multimeter Module Setup Chapter 1 Figure 1 2 Multimeter Measurement Terminals...
Страница 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Страница 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Страница 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...