Multimeter Command Reference
:QUEStionable:ENABle?
STATus:QUEStionable:ENABle?
returns a decimal-weighted number
representing the bits enabled in the Questionable Data/Signal Register's enable
register signifying which bits will set QUE in the Status Byte.
:QUEStionable[:EVENt]?
STATus:QUEStionable[:EVENt]?
returns a decimal-weighted number
representing the bits set in the Questionable Data/Signal Register's event register.
This command clears all bits in the event register when executed.
Figure 3-1. E1412A Status System Register Diagram