150
Multimeter Command Reference
VOLTage[:DC]:RANGe:AUTO
[SENSe:]VOLTage[:DC]:RANGe:AUTO <
mode
>
enables or disables the autorange
function for DC voltage measurements.
Parameters
Comments
•
You can substitute decimal values for the
OFF
(“
0
”) and
ON
(“
1
”) parameters.
•
When autoranging is
ON
, the multimeter samples the input before each
measurement and selects the appropriate range.
•
If you explicitly select a range using
VOLTage:DC:RANGe
, autoranging is
turned
OFF
.
•
In order to specify an aperture time of 10
s, you must select a fixed range
(e.g.,
VOLT:DC:RANG:AUTO OFF
).
•
Related Commands:
CONFigure
,
VOLTage:RANGe
•
*RST
Condition:
VOLT:DC:RANG:AUTO ON
Example
Disable Autoranging
VOLT:DC:RANG:AUTO OFF
Disable autorange.
VOLTage[:DC]:RANGe:AUTO?
[SENSe:]VOLTage[:DC]:RANGe:AUTO?
returns a number to show whether the
autorange mode is enabled or disabled: “
1
” =
ON
, “
0
” =
OFF
. The value is sent to
the output buffer.
Example
Query the Autorange Mode
VOLT:DC:RANG:AUTO OFF
Disable autorange.
VOLT:DC:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
mode
>
boolean
OFF|0|ON|1
none
Содержание Agilent E1412A
Страница 2: ......
Страница 10: ...8 Contents Notes...
Страница 14: ...12 Notes...
Страница 15: ...13 Notes...
Страница 16: ...14 Notes...
Страница 20: ...18 Agilent E1412A Multimeter Module Setup Chapter 1 Figure 1 2 Multimeter Measurement Terminals...
Страница 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Страница 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Страница 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...