136
Multimeter Command Reference
PERiod:APERture
[SENSe:]PERiod:APERture <
time
>|MIN|MAX
sets the integration time in
seconds. Values for time are rounded up to the nearest aperture time shown in the
following table.
Parameters
Comments
•
MINimum
sets the aperture time to 10ms.
MAXimum
sets the aperture time to
1 second.
•
The fastest aperture time available when autoranging is 100ms. In order to
specify an aperture time of 10ms, you must select a fixed range.
•
Setting the aperture time also sets the resolution. Aperture time of 0.01 sets
resolution at 4½-digits, 0.1 sets 5½-digits and 1 sets 6½-digits.
•
*RST
Condition:
0.1 (100ms)
Example
Set the Aperture Time
PER:APER 1E-2
Aperture time is 10 ms.
PERiod:APERture?
[SENSe:]PERiod:APERture? [MIN|MAX]
returns one of the following numbers to
the output buffer:
•
The present aperture time in seconds if MIN or MAX is not specified.
•
The minimum aperture time available (10 ms) if MIN is specified.
•
The maximum aperture time available (100 ms) if MAX is specified.
Example
Query the Aperture Time
PER:APER MIN
Aperture time is 10 ms.
PER:APER?
Query multimeter to return aperture time.
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
time
>
numeric
10ms|100ms|MIN|MAX
seconds
Содержание Agilent E1412A
Страница 2: ......
Страница 10: ...8 Contents Notes...
Страница 14: ...12 Notes...
Страница 15: ...13 Notes...
Страница 16: ...14 Notes...
Страница 20: ...18 Agilent E1412A Multimeter Module Setup Chapter 1 Figure 1 2 Multimeter Measurement Terminals...
Страница 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Страница 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Страница 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...