Technical Information
Vent Accessories
17
TV 551/701 SEM Navigator User Manual / 87-900-958-01(G)
255/
276
The vent device has adjustable delay time (up to 36 min.). This value
can be used only with rack controller.
To install the vent device unscrew and remove the threaded plug (see
the following figure).
Figure 43
Screw the flange mod. 969-9108 on the pump, taking care of the o-
ring right position. Assemble the seal ring and lock the vent device in
position using the KF klamp.
Figure 44