17
Technical Information
J1 - In-Out
224/276
TV 551/701 SEM Navigator User Manual / 87-900-958-01(G
)
PIN N.
SIGNAL NAME INPUT/OUTPUT
12 SPARE
13 FAULT
OUTPUT
O
14 PROGRAMMABLE
ANALOG
SIGNAL (+)
O
15 GROUND
PROGRAMMABLE ANALOG SIGNAL (-)
O
Signal Description
START/STOP
: input signal to start or stop the pump. With the
supplied cover connector the START/STOP (+) signal is connected to
the +24 Vdc pin and the START/STOP (-) signal to the GROUND pin:
in this condition the pump automatically starts as soon as the
controller recognises the input supply ("Plug & Pump").
INTERLOCK
: input signal to control the pump rotation. With the
supplied cover connector the INTERLOCK (+) signal is connected to
the +24 Vdc pin and the INTERLOCK (-)signal to the GROUND pin.
SPEED SETTING
: PWM input signal to set the pump speed. The
PWM signal characteristics must be the following:
frequency: 100 Hz +/-20 %
amplitude: 0 - 24 V max
duty cycle range: from 25 % to 75
%
corresponding to a pump speed from 272 Hz (25 %) to 714 Hz
(75 %) linearly.