6-10
Programmer’s Guide
Trace Data Transfers
Transferring Data with IBASIC
Transferring Data with IBASIC
If you are using IBASIC, your IBASIC program can avoid the overhead of
using
OUTPUT
and
ENTER
to transfer trace data, and instead use the
analyzer's built-in high-speed subprograms. These built-in subroutines
let you quickly move data between the analyzer's measurement arrays
and your program's data arrays. For example, to read the analyzer's
formatted data array, use the following:
10 DIM Fmt(1:201)
20 INTEGER Chan
30 LOADSUB Read_fdata FROM "XFER:MEM 0,0"
40 Chan=1
50 Read_fdata(Chan,Fmt(*))
Refer to the HP Instrument BASIC User's Handbook for more details.
The table below compares the speed of IBASIC using high-speed transfer
subroutines with that of a fast external controller using the SCPI
TRACE:DATA? CH1FDATA
query.
Table 6-3
High-Speed Trace Transfer Times
Number of Trace Points
Controller Using Binary
TRACE:DATA?
(ms)
IBASIC Using
Read_fdata
(ms)
51
21
7
201
23
10
401
30
13
1601
82
32
Содержание 8712ES
Страница 11: ...1 1 1 Introduction to GPIB Programming ...
Страница 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Страница 36: ...3 1 3 Passing Control ...
Страница 39: ...4 1 4 Data Types and Encoding ...
Страница 46: ...5 1 5 Using Status Registers ...
Страница 71: ...6 1 6 Trace Data Transfers ...
Страница 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Страница 99: ...7 1 7 Using Graphics ...
Страница 105: ...8 1 8 Front Panel Keycodes ...
Страница 111: ...9 1 9 Introduction to SCPI ...
Страница 129: ...10 1 10 Menu Map with SCPI Commands ...
Страница 268: ...12 1 12 SCPI Conformance Information ...
Страница 290: ...13 1 13 SCPI Error Messages ...