Technical Information
RS 232/RS 485 Communication Description
17
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01
(B)
267/294
N.
Read/Write Datatype Description
Admitted Values
202
R
N
Pump power in W (pump
current x pump voltage duty
cycle)
203
R
N
Driving frequency in Hz
204
R
N
Pump temperature in °C
0 to 70
205
R
N
Pump status
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
206
R
N
Error code
Bit description:
see the
following figure
300
R
N
Cycle time in minutes (zeroed
by the reset command)
0 to 999999
301
R
N
Cycle number (zeroed by the
reset command)
0 to 9999
302
R
N
Pump life in hours (zeroed by
the reset command)
0 to 999999
320
to
399
Reserved to Agilent service
400
R
A
CRC EPROM (QE)
QE5XXXX
(where “XXXX”
are variable)
402
R
A
CRC Param. (PA)
PA5XXXX
(where “XXXX”
are variable)
500
Reserved to Agilent service
503
R/W
N
RS 485 address
0 to 31
(default = 0)
504
R/W
L
Serial type select
0 = RS 232
1 = RS 485
(default = 0)
Содержание 8698961R001
Страница 3: ...Turbo V 1K G Retrofit Turbo V 1K G Retrofit User Manual 87 900 996 01 B 3 294 Turbo V 1K G Retrofit...
Страница 4: ...Turbo V 1K G Retrofit 4 294 Turbo V 1K G Retrofit User Manual 87 900 996 01 B...
Страница 14: ...Contents 14 294 Turbo V 1K G Retrofit User Manual 87 900 996 01 B...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......