Technical Information
RS 232/RS 485 Communication Description
17
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01
(B)
265/294
Window-Meanings
Tab. 7
Window-Meanings
N.
Read/Write Datatype Description
Admitted Values
000
R/W
L
Start/Stop (in remote mode
the window is a read only)
Start = 1
Stop = 0
008
R/W
L
Remote (default) or Serial
configuration
Remote = 1
Serial = 0
(default = 1)
100
R/W
L
Soft Start
(write only in Stop condition)
YES = 1
NO = 0
101
R/W
N
Set Point type
0 = Frequency
1 = Current
2 = Time
(default = 0)
102
R/W
N
Set Point threshold
(expressed in Hz, mA or s)
(default = 582)
103
R/W
N
Set Point delay: time between
the pump start and the set
point check (seconds)
0 to 99999
(default = 0)
104
R/W
L
Set Point signal activation
type: the signal can be "high
level active" or "low level
active"
0 = high level
active
1 = low level
active
(default = 0)
105
R/W
N
Set point hysteresis
(in % of threshold)
0 to 100
(default = 2)
106
R/W
L
Intercooling
0 = NO
1 = YES
107
R/W
L
Active Stop (write only in
stop)
0 = NO
1 = YES
Содержание 8698961R001
Страница 3: ...Turbo V 1K G Retrofit Turbo V 1K G Retrofit User Manual 87 900 996 01 B 3 294 Turbo V 1K G Retrofit...
Страница 4: ...Turbo V 1K G Retrofit 4 294 Turbo V 1K G Retrofit User Manual 87 900 996 01 B...
Страница 14: ...Contents 14 294 Turbo V 1K G Retrofit User Manual 87 900 996 01 B...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......