17 Technical Information
RS 232/RS 485 Communication Description
266/294
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)
N.
Read/Write Datatype Description
Admitted Values
108
R/W
N
Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
109
W
L
Pump life/ cycle time/ cycle
number reset
To reset write
‘1’
110
R/W
L
Interlock type (default = 1)
Impulse = 0
Continuous = 1
111
R/W
L
Analog output type: output
voltage signal proportional to
frequency or power
0 = frequency
1 = power
(default = 0)
120
R/W
N
Rotational frequency setting
(Hz)
500 to 761
(default = 761)
121
R/W
N
Maximum rotational
frequency in Hz (active only in
Stop condition)
500 to 761
(default = 761)
122
R/W
L
Set vent valve on/off
(on = closed)
On = 1
Off = 0
(default = 1)
123
124
Reserved to Agilent service
125
R/W
L
Set the vent valve operation
Automatic = 0
(see note 1.)
On command =
1 (see note 2.)
126
R/W
N
Vent valve opening delay
(expressed in 0.2 sec)
0 to 65535
(corresponding
to 0 to 13107
sec)
130
Reserved to Agilent service
200
R
N
Pump current in mA dc
201
R
N
Pump voltage in Vdc
Содержание 8698961R001
Страница 3: ...Turbo V 1K G Retrofit Turbo V 1K G Retrofit User Manual 87 900 996 01 B 3 294 Turbo V 1K G Retrofit...
Страница 4: ...Turbo V 1K G Retrofit 4 294 Turbo V 1K G Retrofit User Manual 87 900 996 01 B...
Страница 14: ...Contents 14 294 Turbo V 1K G Retrofit User Manual 87 900 996 01 B...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......