Performance Tests
Performance Test Instructions
68
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
Agilent 81610A/11A/12A/13A/14A:
Relative Uncertainty of Return Loss and
Dynamic Range
The maximum return loss range at the Return Loss Meter is
significantly affected by all parasitic reflections and backscatter
returned to the optical output of the Return Loss Meter.
To minimize these effects, it is essential to use clean connectors, and
to minimize the fiber lengths and the number of optical connections
used in the measurement setup.
1
Make sure that all connectors are clean.
2
Setup the equipment as shown in Figure 39.
3
Press [Preset] on the mainframe.
Figure 39
Relative Uncertainty of Return Loss - Calibration Setup
N O T E
To ensure traceability, use the 81610CC Reference Cable for
calibration measurements. Do not use the 81610CC Reference Cable
for measurements on a Device Under Test. Instead, use a measurement
patchcord.
C A U T I O N
It is important to maintain the quality of the straight connector end of
the 81610CC Reference Cable. Never add another connector to the
straight end of the 81610CC Reference Cable, since a physical
connection is made. When the straight end of the 81610CC Reference
Cable is connected to the 81634A Power Sensor module, no physical
connection is made, so there should be no degradation of connector
quality.
4
Zero the return loss meter as well as the power meter.
5
Enable the internal laser source of the return loss module and allow
20min for it to stabilize.
Mainframe 8163A w/ 8161xA RTL
and 81634A Power Sensor
8163A
8161xA
in
out
81634A
81610CC
Содержание 81610A
Страница 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Страница 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Страница 51: ...Accessories Modules and Options 48 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Страница 57: ...Specifications Definition of Terms 54 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Страница 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Страница 113: ...Cleaning Information Other Cleaning Hints 110 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 115: ...Cleaning Information Other Cleaning Hints 112 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 116: ......
Страница 117: ... Agilent Technologies GmbH 2000 2002 Printed in Germany February 20 2002 81610 90A15 81610 90A15 www agilent com ...