How to clean optical devices which are sensitive to mechanical stress and pressure
Cleaning Information
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
105
Preferred Procedure
Use the following procedure on most occasions.
1
Blow away any dust or dirt with compressed air.
If this is not sufficient, then
2
Clean the interface by rubbing a new, dry cotton swab over the
surface using a small circular movement.
3
Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the interface, and
using the preferred procedure is not sufficient. Using isopropyl
alcohol should be your last choice for recessed lens interfaces because
of the difficulty of cleaning out any dirt that is washed to the edge of
the interface:
1
Moisten a new cotton swab with isopropyl alcohol.
2
Clean the interface by rubbing the cotton swab over the surface
using a small circular movement.
3
Take a new, dry soft tissue and remove the alcohol, dissolved
sediment and dust, by rubbing gently over the surface using a small
circular movement.
4
Blow away any remaining lint with compressed air.
How to clean optical devices
which are sensitive to
mechanical stress and pressure
Some optical devices, such as the Agilent 81000BR Reference
Reflector, which has a gold plated surface, are very sensitive to
mechanical stress or pressure. Do not use cotton swabs, soft tissues or
other mechanical cleaning tools, as these can scratch or destroy the
surface.
Preferred Procedure
Use the following procedure on most occasions.
1
Blow away any dust or dirt with compressed air.
Procedure for Stubborn Dirt
To clean devices that are extremely sensitive to mechanical stress or
pressure you can also use an optical clean polymer film. This
Содержание 81610A
Страница 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Страница 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Страница 51: ...Accessories Modules and Options 48 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Страница 57: ...Specifications Definition of Terms 54 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Страница 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Страница 113: ...Cleaning Information Other Cleaning Hints 110 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 115: ...Cleaning Information Other Cleaning Hints 112 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 116: ......
Страница 117: ... Agilent Technologies GmbH 2000 2002 Printed in Germany February 20 2002 81610 90A15 81610 90A15 www agilent com ...