Specifications
Return Loss Module Specifications with Internal Source
58
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
Table 5
Return Loss Module Specifications with Internal Source
Agilent 81611A
Agilent 81612A
Agilent 81613A
Agilent 81614A
Source
Fabry-Perot Laser (internal)
Output Power
typ. -4dBm
Center wavelength
[1]
1310 nm ± 20 nm typ. 1550 nm ± 20 nm typ.
1310/1550 nm
± 20 nm typ.
1550/1625 nm
± 20 nm typ.
Sensor element
InGaAs
Fiber type
Standard single-mode 9/125 µm
Dynamic Range
[1]
75 dB
Relative uncertainty of Re-
turn Loss (RL)
User Calibration
[2]
Plug and Play
[3]
RL
≤
55 dB
< ±0.5 dB (typ. < ±0.3 dB)
typ. < ±0.6 dB
RL
≤
60 dB
< ±0.6 dB (typ. < ±0.4 dB)
typ. < ±1.5 dB
RL
≤
65 dB
< ±0.8 dB (typ. < ±0.5 dB)
RL
≤
70 dB
< ±1.9 dB (typ. < ±0.8 dB)
[4]
RL
≤
75 dB
typ. < ±2.0 dB
[4]
Total uncertainty
add ±0.2 dB
add typ. ±0.2 dB
Dimensions (H x W x D)
75 mm x 32 mm x 335 mm (2.8” x 1.3” x 13.2”)
Weight
1 kg
Recalibration period
2 years
Operating temperature
10 to 40ºC
Humidy
Non-condensing
Warm-up time
[5]
20 minutes
[1]
At 25ºC constant temperature,
coherence control on
warm-up time after laser turn on >5 min.
[2]
Averaging Time 1 s,
calibration prior to measurement,
constant temperature,
coherence control on,
warm-up time after laser turn on >5 min
length of measurement patchcord
≥
2m,
angled connector in optimal optical condition
Reference Cable 81610CC used for total uncertainty.
[3]
Factory defaults are set to (no user calibration necessary):
length of measurement patchcord
≤ 2
m,
return loss of connectors
≥
70 dB
[4]
For measurements performed immediately after calibration
[5]
Warm-up time 60 minutes if not previously stored at the same temperature.
Содержание 81610A
Страница 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Страница 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Страница 51: ...Accessories Modules and Options 48 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Страница 57: ...Specifications Definition of Terms 54 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Страница 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Страница 113: ...Cleaning Information Other Cleaning Hints 110 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 115: ...Cleaning Information Other Cleaning Hints 112 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 116: ......
Страница 117: ... Agilent Technologies GmbH 2000 2002 Printed in Germany February 20 2002 81610 90A15 81610 90A15 www agilent com ...