Making a Return Loss Measurement
Getting Started with Return Loss
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
33
– [
Meas
], the power currently measured by the Return loss
Module’s internal power sensor in dBm
Figure 26
The Calibration Parameters Screens - Return Loss Diode
2
Press [Next]. The
Calibration parameters
screen for the
Monitor
Diode
appears, as displayed in Figure 27. This screen shows current
value for the following quantities:
– [
Para
], the parasitic power value measured by the Return Loss
module’s monitor diode in dBm. This value is determined by the
termination calibration, or you can use the default setting held in
the factory calibration.
– [
Ref
], the power measured by the Return Loss module’s monitor
diode during the reflectance calibration in dBm
Содержание 81610A
Страница 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Страница 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Страница 51: ...Accessories Modules and Options 48 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Страница 57: ...Specifications Definition of Terms 54 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Страница 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Страница 113: ...Cleaning Information Other Cleaning Hints 110 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 115: ...Cleaning Information Other Cleaning Hints 112 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 116: ......
Страница 117: ... Agilent Technologies GmbH 2000 2002 Printed in Germany February 20 2002 81610 90A15 81610 90A15 www agilent com ...