38
Hardware Installation
Installation
5
Connect the MSD power cord to an appropriate AC outlet.
6
Loosen the sideplate thumbscrews completely (shown in
) and open the analyzer.
7
Remove the interface column nut and blank ferrule from the
end of the GC/MSD interface.
8
Install the column in the GC/MSD interface. See the 5973N or
5973 inert MSD Hardware Manual for more information.
Figure 8
Loosen sideplate thumbscrews
Rear sideplate thumbscrew
(loosen completely)
C A U T I O N
Opening the analyzer with either sideplate thumbscrew tight will
irreparably damage the baseplate. Damage of this type is not covered
by the warranty.
Содержание 5973N
Страница 1: ...Agilent Technologies 5973N and 5973 inert Mass Selective Detector Hardware Installation ...
Страница 6: ...6 Hardware Installation ...
Страница 20: ...20 Hardware Installation General Information ...
Страница 59: ......
Страница 60: ...G2589 90073 Agilent Technologies Agilent Technologies Inc Printed in USA August 2003 ...