30
Hardware Installation
Installation
Preparing the MSD for Installation
Materials needed:
5973N or 5973 inert MSD Hardware Manual
Tygon tubing or hose for pump exhaust (11mm id)
Wrench, 1/4-inch
×
5/16-inch open end
5973 Series MSD Shipping Kit
5973 inert MSD CI Shipping Kit
1
Unpack the MSD and foreline pump, and place them on the
bench near the GC.
2
Remove the analyzer cover of the MSD. See the 5973N or
5973 inert MSD Hardware Manual for more information.
3
Equalize the pressure in the MSD by turning the vent valve
knob counterclockwise 1/2 turn.
The MSD is not under vacuum. It has been backfilled with
clean, dry nitrogen for shipping.
4
Add PFTBA (05971-60571) to the EI calibration vial at the
front of the MSD.
See the 5973N or 5973 inert MSD Hardware Manual for more
information.
WA R N I N G
To prevent injury, get lifting assistance
C A U T I O N
The MSD is shipped with empty calibration vials. The MSD cannot be
tuned without calibration fluid.