Installation
Hardware Installation
37
Connecting the MSD and the GC
1
Verify that the GC power is turned off.
2
Connect the two ends of the GC/MSD interface cable.
3
Carefully position the MSD close to the GC as shown in
When positioned properly, the end of the GC/MSD interface
cover will touch the metal of the oven wall, and the end of the
interface will protrude into the GC oven. Note that the end of
the interface on a CI MSD will not protrude quite as far into
the GC oven as that on an EI MSD.
4
Connect a remote cable between the MSD and one of the
remote connectors on the back of the GC.
Figure 7
Connecting the MSD and the GC
6890 Series GC
5973 Series MSD
GC/MSD interface cable (GC end)
GC/MSD interface cable (MSD end)
GC/MSD interface cover