![Agilent Technologies 53150A Скачать руководство пользователя страница 70](http://html1.mh-extra.com/html/agilent-technologies/53150a/53150a_service-manual_2867598070.webp)
Chapter 3 Replacing Assemblies
Removing the Handle and Handle Mounts
Assembly Level Service Guide
3-5
3
Removing the Handle and Handle Mounts
Use the following procedure to remove the carrying handle and handle mounts:
1
Rotate the handle to the vertical position above the Counter, pull the ends
of the handle outward at the mounts to detach it, and set it aside.
Figure 3-2. Removing the Handle
2
Remove the front bumper (see “Removing the Bumpers” on page 3-4).
3
Use a T15 TORX screwdriver to remove the two screws that attach each of
the handle mounts to the Counter, and remove the mounts.
Figure 3-3. Removing the Handle Mounts
NOTE
When reattaching the handle mounts, limit the torque on the TORX
screws to 4 to 5 inch-pounds.
TORX Screws
TORX Screws
Handle Mount
Handle Mount
Counter
Содержание 53150A
Страница 14: ...1 Performance Tests Verifying Specifications...
Страница 47: ...Chapter 1 Performance Tests Performance Test Record Page 4 of 4 1 34 Assembly Level Service Guide 1...
Страница 48: ...2 Service...
Страница 66: ...3 Replacing Assemblies Disassembly and Reassembly...
Страница 83: ...Chapter 3 Replacing Assemblies Replacing the Power Supply 3 18 Assembly Level Service Guide 3...
Страница 84: ...4 Replaceable Parts...
Страница 95: ...Chapter 4 Replaceable Parts Parts Identification 4 12 Assembly Level Service Guide 4...
Страница 96: ...5 Backdating...
Страница 98: ...6 Specifications...
Страница 109: ...Index Index 4 Assembly Level Service Guide Index...