![Agilent Technologies 53150A Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/agilent-technologies/53150a/53150a_service-manual_2867598022.webp)
Chapter 1 Performance Tests
Operational Verification
Assembly Level Service Guide
1-9
1
Figure 1-1. External Timebase Test Setup
4
Cycle the
POWER
button to preset the Counter.
5
Press the
Shift
key on the Counter’s front panel.
6
Press the
Menu
key on the Counter’s front panel.
7
If necessary, use the up- and down-arrow keys to cycle the display until it
reads
REF OSC> INT
.
8
Press the right-arrow key once.
INT
begins to flash off and on.
9
Press the up- or down-arrow key once.
INT
changes to
EXT
.
10
Press the
Enter
key.
11
Press the
Chan Select
key once to select the
CHANNEL 1
input connector.
Agilent
53150A/51A/52A
Counter
Agilent 3325A/B Synthesizer
Agilent 3325A/B Synthesizer
To Reference 10 MHz
Connector
(Rear Panel)
Output
Output
Channel 1
Содержание 53150A
Страница 14: ...1 Performance Tests Verifying Specifications...
Страница 47: ...Chapter 1 Performance Tests Performance Test Record Page 4 of 4 1 34 Assembly Level Service Guide 1...
Страница 48: ...2 Service...
Страница 66: ...3 Replacing Assemblies Disassembly and Reassembly...
Страница 83: ...Chapter 3 Replacing Assemblies Replacing the Power Supply 3 18 Assembly Level Service Guide 3...
Страница 84: ...4 Replaceable Parts...
Страница 95: ...Chapter 4 Replaceable Parts Parts Identification 4 12 Assembly Level Service Guide 4...
Страница 96: ...5 Backdating...
Страница 98: ...6 Specifications...
Страница 109: ...Index Index 4 Assembly Level Service Guide Index...