![Agilent Technologies 53150A Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/agilent-technologies/53150a/53150a_service-manual_2867598039.webp)
Chapter 1 Performance Tests
Test 4: Channel 2 Power Measurement
1-26
Assembly Level Service Guide
1
5
Connect the Agilent 11900A 2.4 mm (m) to 2.4 mm (m) Adapter to the
Counter’s
CHANNEL 2
input connector.
6
Connect the unused output of the Microwave Splitter to the Adapter on
the Counter’s
CHANNEL 2
input connector.
Figure 1-6. 50 MHz to 20 GHz (–20 dBm) Power Measurement Test Setup
7
Verify that the Counter is still set to
CHANNEL 2
(if necessary, use the
Chan Select
key on the Counter’s front panel to select
CHANNEL 2
).
8
Adjust the power output of the Synthesizer so that the power output of the
Microwave Amplifier is –20 dBm.
9
Set the Synthesizer to the first frequency value listed for Test 4a in the
Performance Test Record on page 1-32 (50 MHz).
10
Record the actual power reading in the Performance Test Record.
Agilent
11667C Power Splitter
Agilent
83051A
Microwave Amplifier
Agilent
8487A/D Power Sensor
Agilent 83650B Synthesizer
Agilent
11900B Adapter
Agilent
53150A/51A/52A
Counter
Channel 2
RF Output
Input
Agilent
437B Power Meter
Содержание 53150A
Страница 14: ...1 Performance Tests Verifying Specifications...
Страница 47: ...Chapter 1 Performance Tests Performance Test Record Page 4 of 4 1 34 Assembly Level Service Guide 1...
Страница 48: ...2 Service...
Страница 66: ...3 Replacing Assemblies Disassembly and Reassembly...
Страница 83: ...Chapter 3 Replacing Assemblies Replacing the Power Supply 3 18 Assembly Level Service Guide 3...
Страница 84: ...4 Replaceable Parts...
Страница 95: ...Chapter 4 Replaceable Parts Parts Identification 4 12 Assembly Level Service Guide 4...
Страница 96: ...5 Backdating...
Страница 98: ...6 Specifications...
Страница 109: ...Index Index 4 Assembly Level Service Guide Index...