Chapter 1 Performance Tests
Operational Verification
1-10
Assembly Level Service Guide
1
12
Set the first Synthesizer (the one connected to the Counter’s rear-panel
Reference 10MHz
connector) to output a 10 MHz, 1V rms, sine-wave
signal.
13
Set the second Synthesizer (the one connected to the
CHANNEL 1
input) to
output a 2 MHz, 100 mV rms, sine-wave signal.
14
Press and release the Counter’s
Reset/Local
key.
15
Verify that the Counter displays a reading of approximately 2 MHz and
that the
Ext Ref
annunciator is activated.
16
Mark Pass or Fail for Test 4 in the Performance Test Record on page 1-30.
This completes the Operational Verification.
NOTES
•
If you intend to perform the Complete Performance Tests next, do not
disconnect the cable from the Counter’s
Reference 10MHz
connector.
This connection is used in the Complete Performance Tests.
•
Do not turn off the Counter if you intend to perform the Complete
Performance tests. Turning the Counter off and back on resets the
reference oscillator (
REF OSC
) selection to internal (
INT
) and the
channel selection to
CHANNEL 2
.
Содержание 53150A
Страница 14: ...1 Performance Tests Verifying Specifications...
Страница 47: ...Chapter 1 Performance Tests Performance Test Record Page 4 of 4 1 34 Assembly Level Service Guide 1...
Страница 48: ...2 Service...
Страница 66: ...3 Replacing Assemblies Disassembly and Reassembly...
Страница 83: ...Chapter 3 Replacing Assemblies Replacing the Power Supply 3 18 Assembly Level Service Guide 3...
Страница 84: ...4 Replaceable Parts...
Страница 95: ...Chapter 4 Replaceable Parts Parts Identification 4 12 Assembly Level Service Guide 4...
Страница 96: ...5 Backdating...
Страница 98: ...6 Specifications...
Страница 109: ...Index Index 4 Assembly Level Service Guide Index...