Chapter 2 Service
Calibration Procedure
Assembly Level Service Guide
2-7
2
Figure 2-1. Calibrating the Counter
3
Insert a non-metallic slotted adjustment tool through the hole in the
lower-right side of the Counter’s cover and into the slot in the timebase
calibration screw.
4
Turn the calibration screw clockwise to increase the Counter’s reading or
counter-clockwise to decrease the reading until the frequency value
displayed on the Counter’s front panel is exactly
10,000,000 Hz
.
Calibration
Access Hole