4278A
1.
Connect
the
attachment
with
the
cover
to
the
Guarded/Guard
electrode
.
2.
Press
N
NNNNNNNNNNNNNNNNNNN
COMPEN
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OPEN
COMPEN
to
start
the
OPEN
oset
measurement
and
observe
the
display
.
The
message
Open
offset
compen.
completed
will
be
displayed
on
the
Message
Line
(the
bottom
line
of
the
LCD),
and
the
OPEN
admittance
data
will
be
displayed
on
the
Monitor
Line
(the
second
line
from
the
bottom).
3.
Press
4
TRIGGER
5
and
conrm
that
the
measurement
result
is
the
same
(or
almost
the
same)
as
the
OPEN
admittance
data.
4.
Press
N
NNNNNNNNNNNNNNNNNNNNNN
OPEN
ON
,
then
press
4
TRIGGER
5 .
The
measurement
result
is
compensated
for
,
in
accordance
with
the
OPEN
admittance
data,
and
is
close
to
zero
in
value
.
Press
NNNNNNNNNNNNNN
NEXT
.
5.
Remove
the
attachment
from
the
electrode
.
6.
Connect
the
attachment
without
the
cover
to
the
Unguarded
electrode
of
the
16451B.
7.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SHORT
COMPEN
to
start
the
SHORT
correction
measurement,
then
the
message
Short
offset
compen.
completed
will
be
displayed
on
the
Message
Line
,
and
the
SHORT
correction
impedance
data
will
be
displayed
on
the
Monitor
Line
.
8.
Press
N
NNNNNNNNNNNNNNNNNNNNNNNNN
SHORT
ON
and
N
NNNNNNNNNNNNNNNNNNN
return
.
9.
Remove
the
attachment
from
the
electrodes
and
store
it
with
the
cover
in
the
carrying
case
.
Correction
Procedure
C-5
Содержание 16451B
Страница 8: ... C Copyright 2008 Agilent Technologies ...
Страница 10: ......
Страница 20: ......
Страница 28: ......
Страница 34: ......
Страница 46: ......
Страница 56: ...Figure 3 6 Summary of Measurement Methods 3 10 Operation ...
Страница 83: ...because a change of temperature causes mechanical dimensions to change Operation 3 37 ...
Страница 119: ...Figure 4 4 Cable Connection Diagram Service 4 9 ...
Страница 120: ......
Страница 122: ......
Страница 132: ......
Страница 138: ......