7-12 11713 Operating and Service Manual
Replacing Major Assemblies
A5 GPIB Handshake Data Input Assembly Removal/Replacement
A5 GPIB Handshake Data Input Assembly
Removal/Replacement
CAUTION
This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe
workstation. Refer to "Preparing a Static-Safe Work Station" in
Chapter 8.
To Remove the A5 GPIB Assembly
1. Remove the top cover
2. With a strong upward motion, lift the green ring at the top edge of
the A5 GPIB handshake data input assembly.
To Replace the A5 GPIB Assembly
1. Gently insert the board with a vertical motion into the socket
marked A5 GPIB handshake data input assembly. Apply balanced
pressure with your thumbs to push the board into the socket.
2. Replace the top cover