2-8 11713A Operating and Service Manual
Installing the Agilent Technologies 11713A Attenuator/Switch Driver
Table 2-2 Attenuator Cable Connector Pin Numbers and Wire
Color Codes
For the SWITCHES, if pin 3 is grounded (lamp on), pin 4 is connected to +24 V.
Front Panel Pushbuttons
a
ATTENUATOR X, Y, and SWITCHES
Rear Panel connectors ATTEN X or Y, and
Attenuator Cable
Pushbutton
Number
Pushbutton LEDs
Pin Numbers
Attenuator Cable Wire
Color Code
1
Red (24 Vdc)
2
White/Brown (Gnd)
SWITCHE
S
Connect to S9 and
S0 outputs
9
ON
ATTEN X 3(S9-A)
Gray
9
OFF
ATTEN X 4(S9-B0
White/Red
0
ON
ATTEN Y 3 (S0-A)
Grey
0
OFF
ATTEN Y 4 (S0-B)
White/Red
ATTENUATORS
X
Y
1
5
OFF
5
Violet
ON
6
Yellow
2
6
OFF
7
Black
ON
8
Green
3
7
OFF
9
Orange
ON
10
Blue
4
8
OFF
11
Brown
ON
12
White
a.
The ON/OFF status of the pushbutton lamps indicates which cable wire or pin on the
rear panel connector is grounded.For example, if ATTENUATOR X pushbutton 3 is illu-
minated, pin 10 of the ATTEN X connector (blue wire of the cable) is grounded and pin 9
floats at a high impedance.