99
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
5500/5503 R
EFERENCE
MANUAL
6
On the ‘Image quality’ screen press the Down key twice, followed by the
Confirm key to select ‘Dmax’.
The ‘Dmax (maximum density)’ screen for the selected modality is displayed:
All maximum densities are selectable starting from 2.75 until 3.20 in steps of 0.05
units. The current setting for the selected modality is marked with an *.
7
Press the Up/Down arrow keys to select the desired Maximum Density setting
for this modality.
8
Press the Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
The maximum density will be limited to the maximum density of the media.
<Modality name>
*2.80
2.85
2.90
2.95
Содержание Drystar 5500
Страница 1: ...Drystar 5500 5503 Reference manual 2900H EN 20071108 ...
Страница 6: ...6 2900H EN 20071108 DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 176: ...176 2900H EN 20071108 Controlling the Drystar 5500 via a remote PC with browser DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 226: ...226 2900H EN 20071108 Troubleshooting DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 227: ...Equipment information sheet Appendix A ...
Страница 237: ...Drystar media density response data Appendix B ...
Страница 241: ...Quality Control Charts Appendix C ...
Страница 252: ...252 2900H EN 20071108 Quality Control Charts DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 253: ...Remarks for HF emission and immunity Appendix D ...
Страница 259: ...Index Chapter E ...
Страница 265: ...265 2900H EN 20071108 Index DRYSTAR 5500 5503 REFERENCE MANUAL ...
Страница 266: ...0413 Printed in Belgium Published by Agfa HealthCare N V B 2640 Mortsel Belgium 2900H EN 20071108 ...