Registering the DR Detector to an NX Workstation using the
DR 10s DR 14s Registration Tool
1.
Start the NX workstation.
2.
Turn on the detector.
During startup the status indicator is flashing orange. After startup the
status indicator is lit green, indicating the power status.
3.
On the NX workstation, go to the
Main Menu
4.
Click the
Show Desktop
action button.
The Windows desktop is shown.
5.
Click the
DR 10s DR 14s Registration Tool
icon on the desktop.
A dialog is displayed instructing to remove all DR Detectors away from the
IR Data Communication Unit.
Note
:
Other detectors in the direct vicinity may interfere
with the communication to the NX workstation. Make sure
they are out of reach of the IR Data Communication Unit.
6.
Click
OK
.
A dialog is displayed instructing to hold the detector with its IR data port
close to the IR Data Communication Unit.
7.
Hold the detector with its IR data port close to the IR Data Communication
Unit that is connected to the NX workstation.
A dialog is displayed on the NX workstation indicating that the network
settings of the DR Detector are being updated.
Note
:
Do not cover the IR data port of the DR Detector with
your hands. Otherwise, the properties of wireless
communication, such as the throughput and operable distance,
may decrease.
After a short time another dialog is displayed indicating that sharing the
DR Detector has been successful. It may take up to 30 seconds for the
dialog to pop up.
The DR Detector is restarting.
8.
Remove the detector from the IR Data Communication Unit and click
OK
.
9.
In specific configurations, the DR Detector turns to error status directly
after registration to another NX workstation, or after attempting a first
DR 10s | Advanced Operating |
91