WARNING:
Do not hit or drop the equipment. The equipment may be
damaged if it receives a strong jolt, which may result in fire or
electric shock if the equipment is used without being repaired.
WARNING:
Have the patient take a fixed posture and do not let the patient
touch parts unnecessarily. If the patient touches connectors or
switches, it may result in electric shock or malfunction of the
equipment.
WARNING:
To avoid electric shocks and burns caused by use of the wrong
type of fire extinguisher, make sure that the fire extinguisher at
the site has been approved for use on electrical fires.
WARNING:
System unavailability due to hardware or software failure. If the
product is used in critical clinical workflows, a backup system
has to be foreseen.
CAUTION:
Strictly observe all warnings, cautions, notes and safety
markings within this document and on the product.
CAUTION:
All Agfa medical products must be used by trained and qualified
personnel.
CAUTION:
This device is not intended to supply heat to a patient. However,
during normal use, surfaces will become heated due to power
dissipation. Patient contact surfaces will not exceed 48 ºC under
normal useconditions. The operator should monitor and
evaluate how much of the patient’s body area is in contact with
these surfaces and for how long.
CAUTION:
Excessive ambient temperature may impact performance of DR
Detectors and cause permanent damage to the equipment. If
ambient temperature and humidity is outside the range of 15 -
35 °C and 20 - 80% RH, do not operate the system or use air
conditioning. Warranty will be void if it is obvious that
operating conditions are not met.
CAUTION:
Turn OFF the power to each piece of equipment for safety when
not being used.