Installation
Installation and configuration is performed by an Agfa trained and authorized
service engineer. Contact your local support organization for more
information.
On a configuration with multiple DR Detectors of the same type, it is required
to apply labeling to the DR Detector containing a unique nickname for each
DR Detector. The nicknames must be configured on the NX Workstation. The
DR Detector Switch shows which DR Detector is active and shows its status, by
means of the nickname of the DR Detector.
An identical label is attached to the bucky of the X-ray system to identify the
dedicated workspace of each DR Detector.
Environment of Use
The equipment is mainly for use in X-ray exposure rooms, hospital wards and
mobile medical examination vehicles. To use it in other places, consult your
sales representative or local Agfa dealer.
WARNING:
Do not install or store the equipment in any of the locations
listed below. Doing so may result in failure or malfunction,
equipment falling, or fire or injury:
•
Close to facilities where water is used
•
Where it will be exposed to direct sunlight
•
Close to the air outlet of an air-conditioner or ventilation equipment
•
Close to a heat source such as a heater
•
In a dusty environment
•
In a saline or sulfurous environment
•
Where temperature or humidity is high
•
Where there is freezing or condensation
•
In areas prone to vibration
•
On an incline or in an unstable area
Note
:
Do not use the detector near devices generating a strong
magnetic field. Doing so may produce image noise or artifacts.
Note
:
Do not use this equipment in combination with peripherals
such as defibrillators or large electric motors as these may cause
power-supply noise or power supply voltage variations. Doing so
may prevent normal operation of this equipment and peripherals.