Daily inspection
WARNING:
For safety reasons, be sure to turn OFF the power to each piece
of equipment before performing the following. Otherwise, an
electric shock may result.
1.
Check the cables
a) Ensure that cables are not damaged and cable jackets are not torn.
b) Ensure that the power cord plugs are securely connected to both the
equipment AC inlet and the AC outlet.
2.
Check the detector
a) Ensure that there are no loose screws or breaks.
b) Ensure that there is no dust or foreign matter on the battery bay
connector.
c) Ensure that there are no breaks or short-circuits in the battery bay
connector.
3.
Start the NX workstation and perform a test exposure.
DR 10s | Introduction to the DR Detector |
51
0351A EN 20190319 1349 *