Safety Directions for the X-Ray System
WARNING:
Ionizing radiation can lead to radiation injuries if handled
incorrectly. When radiation is applied, the required protective
measures must be complied with.
WARNING:
The operator must take precautions to protect himself against
dangerous X-ray exposure when using the DR detector in the X-
ray beam path of an X-ray source.
WARNING:
The DR Detector is not intended to be used as a primary barrier
to X-rays. The user is responsible for ensuring the safety of the
operator, bystanders, and the subjects being radiographed.
WARNING:
Repeated exposures to a patient with high doses can lead to
deterministic effects. Therefore exposure settings shall be
selected carefully and in accordance to the patient and the
object to expose and balanced in such a way that patient dose is
as low as possible while image quality is usable for diagnosis.
WARNING:
Using an inappropriate SID can result in a suboptimal or
unusable image. Too short SID causes warped image.
WARNING:
The use of a filter that is not suitable for the type of examination
can affect the image quality or the amount of radiation.
WARNING:
A wrongly inserted DR detector (e.g. upside down) will result in
an unusable image.
WARNING:
Software failure causing delay in synchronization between DR
detector and generator can result in an unusuable image.
CAUTION:
Avoid unnecessary dose by checking before exposure if the DR
Detector Switch displays the name of the DR Detector that is
being used and if the status of the DR Detector is ready for
exposure.