Loading of the bucky in the wall stand
To load the bucky with a detector:
1
2
5
3
3
44
Figure 84: Loading of the bucky in the wall stand
1.
Open the bucky drawer completely by pulling the front handle.
2.
Adjust the detachable clamp by attaching it to the slit that corresponds to
the bottom alignment of the detector.
• Detector height 17 inch (43 cm): no clamp needed
• Detector height 14 inch (35 cm): use the lower position of the clamp
for centering the detector in the bucky, and the higher position of the
clamp for aligning the detector to the upper edge of the bucky.
3.
Let the detector rest on the clamp, while keeping it in position with one
hand.
4.
Close the clamps to fix the detector position.
CAUTION:
Make sure your fingers are not between the clamping
mechanism and the detector. The clamping mechanism may
hurt your fingers, therefore take special care.
5.
Lock the clamps by turning the lock lever clockwise.
Figure 85: Lock icon