Remarks for HF-emission and immunity
It is hereby certified that the device has interference suppression according to
the EN 55011 Class A as well as the FCC Rules CFR 47 Part 15 Class A.
This device was tested for a normal hospital environment as described above.
The user of the device should ensure that it is used in such an environment.
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment
generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user
will be required to correct the interference at his own expense.
WARNING:
This device is intended for use by healthcare professionals only.
This device may cause radio interference or may disrupt the
operation of nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or relocating the
device or shielding the location.
WARNING:
The HF-emission and immunity can be influenced by connected
data cables depending on length and the manner of installation.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
RF Emission
Measure-
ments
Agree-
ment
Electromagnetic Environment Guidelines
High frequen-
cy RF emis-
sions in ac-
cordance with
CISPR 11
Group 1 The device uses high frequency energy exclusive-
ly for its internal functions. For this reason, its
high frequency RF emission is very low and it is
improbable that neighboring electronic equip-
ment will be disrupted.
High frequen-
cy RF emis-
sions in ac-
cordance with
CISPR 11
Class A
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a resi-
dential environment (for which CISPR 11 class B
is normally required) this equipment might not