Starting and stopping the generator
The generator is switched on and off by the power buttons on the X-ray
generator mini console.
Press the Power ON button on the X-ray generator control box to switch
on the generator.
Press the Power OFF button on the X-ray generator control box to
switch off the generator.
To restart the system, wait at least 10 seconds between stopping and starting
the generator, to allow all components to properly shut down.
Following warning is printed on the X-ray generator mini console in English:
WARNING:
This x-ray unit may be dangerous to patient and operator unless
safe exposure factors, operating instructions and maintenance
schedules are observed.
162
| Valory (system with floor mounted column) | X-Ray Generator Mini Console
0426A EN 20220804 1858