Centering the tube head on the table
To move the X-ray tube head to the center of the DR detector in the bucky of
the radiographic table and to keep the bucky centered while moving the tube
head left and right:
1.
On the X-ray tube head, while pressing the transversal movement button,
move the X-ray tube towards the center of the radiographic table in
transversal direction.
Figure 55: Transversal movement button
There is a stop position for centering on the radiographic table.
2.
While pressing the longitudinal movement button, move the X-ray tube to
the required position.
Figure 56: Longitudinal movement button
3.
If the exposure is oblique, press and hold the tilting button, to adjust the
position of the X-ray tube.
Figure 57: Tilting button
4.
Press the automatic centering and position tracking button.
Figure 58: Position tracking button
The position tracking icon is displayed.
Figure 59: Position tracking icon
The bucky is moved automatically to align it to the X-ray tube.
Valory (system with floor mounted column) | Positioning the X-Ray Tube |
131
0426A EN 20220804 1858