9.5
Overpressure safety device
If, for operational reasons, the range must be smaller than the maximum
operating pressure, install an upstream overpressure safety device to
help protect the product from damage.
In the case of a pressure surge, the overpressure safety device closes imme-
diately, in the case of a slow pressure increase, it closes gradually. Therefore,
the closing pressure to be set depends on the values along the progression.
9.6
Pressure gauge with maximum pointer
Since Bourdon tubes have a relatively small resetting force, you must
account for their influence when installing a maximum pointer. Maximum
pointers may only be used with pressure gauge types D4 and D8 at a mini-
mum range of 6 bar.
WARNING
OVERPRESSURE SAFETY DEVICE WITHOUT EFFECT
Highly viscous and polluted media can render the overpressure safety device
ineffective.
•
Do not use highly viscous or polluted media if you have installed an overpres-
sure safety device.
Failure to follow these instructions can result in serious injury or equip-
ment damage.
Содержание KP EX Series
Страница 41: ...41 Anhang DE RF EX KP EX PF EX 16 Anhang 16 1 Zulassungsunterlagen ...
Страница 42: ...42 Anhang DE RF EX KP EX PF EX 16 2 EU Konformitätserklärung ...
Страница 43: ...43 Anhang DE RF EX KP EX PF EX ...
Страница 44: ...44 Anhang DE RF EX KP EX PF EX 16 3 Informationen zur Druckgeräterichtlinie ...
Страница 84: ...40 Appendix EN RF EX KP EX PF EX 16 Appendix 16 1 Approval documents ...
Страница 85: ...41 Appendix EN RF EX KP EX PF EX 16 2 EU Declaration of Conformity ...
Страница 86: ...42 Appendix EN RF EX KP EX PF EX ...
Страница 87: ...43 Appendix EN RF EX KP EX PF EX 16 3 Information on the Pressure Equipment Directive ...