![Afriso EURO-INDEX D0 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/afriso-euro-index/d0/d0_operating-instructions-manual_2864879019.webp)
Technical specifications
Bourdon tube, capsule, diaphragm,
spring diaphragm pressure gauges
19
Table 5: Technical specifications diaphragm pressure gauges
Parameter
Value
General specifications
Measuring ranges
0/10 mbar to 0/25 mbar
Temperature perfor-
mance
Rising temperature approx. +0.08 %/K
Falling temperature approx. -0.08 %/K
(indication error when the temperature
of the measuring system deviates from
the normal temperature of 20 °C, with
reference to the full scale value)
Operating frequency in
hazardous areas (EX
areas).
Max. 0.1 Hz
Operating temperature range
Ambient
-20 °C to +60 °C
Fluid
Max. +60 °C for liquid filled devices
Max. +100 °C for non-filled devices
Storage
-40 °C to +70 °C
Application area with
static load
Up to full scale value