Aeroneb
®
Solo System Instruction Manual
Immunity test IEC/EN 60601
Test level
Compliance
level
Electromagnetic
environment –
guidance
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % Ut
(>95 % dip
in Ut) for
0.5 cycle
40 % Ut
(60 % dip
in Ut) for
5 cycles
70 % Ut
(30 % dip
in Ut) for
25 cycles
<5 % Ut
(>95 % dip in
Ut) for 5 sec
<5 % Ut
(>95 % dip
in Ut) for
0.5 cycle
40 % Ut
(60 % dip
in Ut) for
5 cycles
70 % Ut
(30 % dip
in Ut) for
25 cycles
<5 % Ut
(>95 % dip in
Ut) for 5 sec
Mains power
quality should be
that of a typical
commercial
or hospital
environment. If
the user of the
Aeroneb Solo
nebulizer system
requires continued
operation during
power mains
interruption, it is
recommended that
the Aeroneb Solo
nebulizer system
be powered from
an uninterruptible
power supply or
battery.
Power
frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Note: Ut is the A.C. mains voltage prior to application of the
test level.
Содержание Solo System
Страница 1: ...Rx Only US 8 30 mins Symbols for Page 8 and 9 Instruction Manual...
Страница 2: ......
Страница 3: ...Part Number AG AS3050 EN Rev J 2011 Aerogen Ltd Instruction Manual Aeroneb Solo System...
Страница 4: ...This page has been intentionally left blank...
Страница 46: ...This page has been intentionally left blank...
Страница 47: ...Aeroneb Solo System Instruction Manual 41 Appendix 1 EMC tables...
Страница 48: ...This page has been intentionally left blank...
Страница 55: ......