Guidance and manufacturer’s declaration – electromagnetic
immunity
The Aeroneb Solo nebulizer system is intended for use in the
electromagnetic environment specified below. The customer or the
user of the Aeroneb Solo nebulizer system should assure that it is
used in such an environment.
Immunity test IEC/EN 60601
Test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Electrical fast
Transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Surge
IEC 61000-4-5
±1 kV
differential
mode
±2 kV
common
mode
±1 kV
differential
mode
±2 kV
common
mode
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Содержание Solo System
Страница 1: ...Rx Only US 8 30 mins Symbols for Page 8 and 9 Instruction Manual...
Страница 2: ......
Страница 3: ...Part Number AG AS3050 EN Rev J 2011 Aerogen Ltd Instruction Manual Aeroneb Solo System...
Страница 4: ...This page has been intentionally left blank...
Страница 46: ...This page has been intentionally left blank...
Страница 47: ...Aeroneb Solo System Instruction Manual 41 Appendix 1 EMC tables...
Страница 48: ...This page has been intentionally left blank...
Страница 55: ......