Solo System Instruction Manual
Cautions
Refer to Figure 6.
The recommended syringe pump software setting with the
Aerogen syringe is typically the “BD Plastipak” setting. This
must be validated locally before use. Refer to pump manual or
manufacturer for guidance. These pumps may also be used in
accordance with local hospital or ward policies.
Ensure that the tethered silicone plug (7) is attached to the
Aeroneb
®
Solo (3) when connecting tube set.
Ensure that the tubing (6) is safely orientated to prevent a trip
hazard.
Rising level of medication in the reservoir may occur if the
Aeroneb Solo nebulizer is turned off while the feed system is
still on or the nebulizer is not in its recommended orientation.
The level of the medication in the reservoir of the Aeroneb Solo
nebulizer should be periodically monitored to ensure that the fill
rate of medication does not exceed the output rate of the
nebulizer. A rising level of medication in the reservoir indicates
that the fill rate is exceeding the output rate of the nebulizer.
Do not clean or sterilize the device.
Replace the tube set and syringe when changing the type of
medication.
Содержание Solo System
Страница 1: ...Rx Only US 8 30 mins Symbols for Page 8 and 9 Instruction Manual...
Страница 2: ......
Страница 3: ...Part Number AG AS3050 EN Rev J 2011 Aerogen Ltd Instruction Manual Aeroneb Solo System...
Страница 4: ...This page has been intentionally left blank...
Страница 46: ...This page has been intentionally left blank...
Страница 47: ...Aeroneb Solo System Instruction Manual 41 Appendix 1 EMC tables...
Страница 48: ...This page has been intentionally left blank...
Страница 55: ......